SPECIFICATIONS
Conductivity Range | 0 . 0 0 t o 1 . 9 9 9 ,19.99,199.9,1999 s, 19.99, 200.0 ms |
Conductivity Accuracy | 1 % F . S |
Temperature Range | 0 t o 1 0 0 C |
Temperature Accuracy | + 1 C |
Temperature Coefficient | 0 to 3.9% per C |
Calibration Points | 1 point per range |
Calibration Solution | 10 to 17 S, 100 to 170 s 1000 to 1700 S, 10 to 17 mS |
Cell Constant | 0.1, 1.0, 10 adjustable |
Temperature Compensation | 0 to 50C, Manual or Automatic |
Power Requirements | DC 9V, using AC Adapter 220VAC, 50Hz |
Dimensions | 2 1 0 ( L ) m m X 205 ( W ) mm X 65 ( H ) mm |
Weight | 1.75 kg |
STANDARD FEATURES:
AUTOMATIC CALIBRATION
The meter is capable of calibrating up to 5 points using standard solution.
From Setup Menu, choose CAL (Calibration) and press key to set known
conductivity value, immerse conductivity electrode into calibration solution,
press ENTER key and meter is automatically calibrated.
CONDUCTIVITY AUTO-RANGING
WCM Series have auto-range function for measuring conductivity. Upon
immersing conductivity sensor into sample solution, the screen immediately
displays the value in appropriate measurement range for accuracy.
AUTOMATIC TEMPERATURE COMPENSATION
Using the supplied temperature probe, the temperature value is displayed
on the screen. The meter automatically compensates for optimum
accuracy under variable temperature conditions.
ADJUSTABLE TEMPERATURE COEFFICIENT AND CELL CONSTANT
Various Ions may have different temperature coefficients. The meters can
set the temperature coefficient between 0 and 3.9% in degrees
centigrade for optimum accuracy, set three cell constants(O.1,1 and 10)