These Generators are designed to be connected to an existing air supply and can cater for the requirements of one or multiple AB SCIEX LCMS.
TECHNICAL SPECIFICATIONS
Infinity 1031 | Infinity 1032 | Infinity 1033 | Infinity 1034 | |
Curtain Gas (Nitrogen) | 18 l/min | 36 l/min | 54 l/min | 72 l/min |
Source Gas (Dry Air) | 26 l/min | 52 l/min | 78 l/min | 104 l/min |
Exhaust Gas (Dry Air) | 25 l/min | 50 l/min | 75 l/min | 100 l/min |
Max Gas Outlet Pressure | 110 psi | |||
Min Air Inlet Flow Rate | 115 L/min | 240 L/min | 340 L/min | 475 L/min |
Min/ Max Air Inlet Pressure | 8.27 - 10.0 bar / 120 - 145 psi | |||
Max Pressure Drop | 8 psi | |||
Min/ Max Operating Temperature | 5C - 30C / 41F - 86F | |||
Particles | < 0.01 m | |||
Phthalates | None | |||
Electrical Requirements | None | |||
Dimensions (cm/ ins) WxDxH | 25 x 20 x 76 10 x 8 x 30 | 31 x 25 x 73 12 x 10 x 29 | 31 x 25 x 73 12 x 10 x 29 | 25 x 25 x 96 10 x 10 x 37.8 |
Weight (kg/ lbs) | 13 / 29 | 21 / 46 | 27 / 60 | 27 / 60 |